Two labs can sieve the same powder and report different curves without either of them making a mistake. The difference is almost always in three variables that nobody wrote down.
By DONGZHEN · November 19, 2026 · 5-minute read
The aperture series is chosen before the run and the machine is whatever sits on the bench. What remains is charge, time and end point — and those three decide whether the number is reproducible.
Charge is the mass you put on the top sieve. Time is how long the stack is shaken. End point is the criterion that tells you to stop, which is not the same as the clock. A method that specifies only "shake for ten minutes" has fixed one variable and left two to habit.
A sieve separates a thin bed well and a thick bed badly. In a deep bed, the particles near the top never reach the cloth; they are supported by the particles below, and they report as oversize. The result is a curve that drifts coarse as the charge grows, and the drift is larger for fine apertures, where the same mass occupies more volume.
That is why the published batch limits on a test sieve shaker are ceilings rather than targets. On the two standard frames the ceilings are 200 g and 300 g, which works out to 6.4 and 4.2 kg per square metre of cloth. Those numbers are useful as a sanity check: if your method puts far more than that on the stack, the top fractions are suspect.
The practical rule is to use the smallest charge that still leaves enough mass in each fraction to weigh accurately. A 200 g charge split across nine fractions leaves an average of 22 g per fraction, which is comfortable on a balance reading to 0.01 g; a 2 kg charge would be more precise in theory and worse in fact, because the separation itself degrades.
Shaking time has a shape: mass passes quickly at first, then slowly, then barely at all. The defensible end point is a mass criterion, not a duration — keep shaking in increments and stop when the additional mass passing in one increment is a negligible share of the charge.
Write that criterion into the method with the increment length. "Shake until the mass passing in one minute is below the stated share of the charge" travels between labs; "shake for ten minutes" does not, because ten minutes on one machine is a different amount of work than ten minutes on another.
Agitation setting belongs in the same sentence. Amplitude and speed determine how much energy the stack receives, and a method that fixes time but not amplitude is only half specified. On the two published frames, speed is fixed at 1450 r/min and amplitude is adjustable up to 5 mm, so amplitude is the variable that has to be recorded.
Performance is decided by the particles closest to the aperture. Published modelling classifies feed into particles below 0.7 times the aperture, particles between 0.7 and 1.0 times, and particles above it, and finds that the middle group inhibits screening efficiency while the coarse group actually helps by disturbing the bed (Shen et al., 2025).
The practical consequence for a lab is that repeatability is not uniform across materials. A powder whose particles cluster around the cut will give a wider spread between duplicate runs than a powder with a clean gap either side of the aperture — and no amount of care removes that spread, because it comes from the material.
Report the near-size share alongside the curve. It tells the reader how much confidence the number deserves, and it is the first thing a technical reviewer will ask for.
A stretched or damaged cloth biases results in one direction and does nothing visible. The sieve still looks fine, the stack still clamps, and the fractions quietly shift.
Two habits catch it. Inspect the cloth under light before each campaign, looking for distorted openings and loose weave at the frame. And verify the stack periodically against a reference material or a calibrated sieve, so the bias is measured rather than assumed. Neither habit is expensive; both are cheaper than a customer dispute about a certificate.
Until a stated mass criterion is met, not for a fixed time. Shake in increments and stop when the mass passing in one increment falls below the share you wrote into the method.
The smallest charge that leaves weighable fractions. Use the machine's published ceiling as a limit rather than a target — for the standard frames that is 200 g and 300 g, or 6.4 and 4.2 kg per square metre.
Because charge, amplitude, end point or cloth condition differed, and usually more than one of them. Compare methods line by line before comparing results.
Write five lines and the analysis becomes transferable: aperture series, charge in grams, amplitude setting, increment length, and the mass criterion for stopping. Add the verification interval for the cloth.
If the material sits in the difficult range at the fine end, the method may need water rather than a longer shake — that decision is set out in wet sieving versus dry sieving. If the stack is still being chosen, what a test sieve shaker actually holds covers deck count and height, and the CW test sieve shaker range lists both frames with their charge limits.
About DONGZHEN
Xinxiang DONGZHEN Machinery Co., Ltd. (CHENWEI Machinery) is a food processing and packaging machinery manufacturer based in Xinxiang, Henan, China. Its screening range covers rotary vibrating screens, ultrasonic vibrating screens, direct discharge screens, vibration filters, tumbler sieves and test sieve shakers for powders, granules, pastes and slurries in food, pharmaceutical and chemical production. Cutting, welding, polishing, assembly and performance testing are handled in-house, non-standard configurations are available, and machines are exported to Southeast Asia, the Middle East, Africa and South America.